- Contact Person : Mr. Feng Charles
- Company Name : Dandong Aolong Radiative Instrument Co., Ltd.
- Tel : 86-415-3458128
- Fax : 86-415-3458588
- Address : Liaoning,DANDONG,NO.46,FUMIN STREET,LINGANG HIGH-TECH ZONE,DANDONG CITY,CN
- Country/Region : China
- Zip : 118009
Mobile Constant Potential X-ray flaw Detectors(XYD-22510/3 model)
Mobile Constant Potential X-ray flaw Detectors Technological Scheme
(XYD-22510/3 model)
1. Summarization
Mobile industrial X-ray flaw detectors is main NDT equipment,Its main function is happen constant potential x-ray, control and adjust and it,may detect the inside defection of workpiece via radiography .
2. Composing1) Metal ceramic X-ray tube MXR-225/22 1Pcs made by COMET
2) High Voltage cable 225KV 5m 1Pcs made by COMET.
3) High Voltage generator T1016 1Pcs made by ALRC
4) Controller T6000 1Pcs made by ALRC
5) Water cooling system ZLB-4 1Pcs made by ALRC
6) device for the x-ray tube moving and fixing 1Pcs made by ALRC
7) Low Voltage cable and the water pipe 1Set made by ALRC
8) Spare parts 1Set made by ALRC
3. The Main Technical Parameter3.1 Industrial Unipolar metal Ceramic X-ray tube, made in Switzerland Comet
3.2 Tube voltage: 40~225KV, adjust constantly and the step is 1Kv
3.3 Tube current: 0.1~10mA(big focus) 0.1~2.8mA(small focus)
Adjust constantly and the step is 0.1 mA
3.4 The focus size:d=5.5mm(big focus) d=1.0mm(small focus) (EN12543)
3.0×3.0 (big focus) 0.4×0.4mm(small focus)(IEC336)
3.5 Cooling mode: Anode water cooling minimum water flow >4L/min
3.6 Radiation coverage: 40°
3.7 Exposure condition:
Tube voltage: 225KV
Tube current: 10mA
Double foil lead intensifying screen
Focus distance: 600mm
Exposure time: 10min
Developing temperature: 20°C
Industrial X-ray film
3.8 Maximal penetration (Fe):43mm
3.9 Sensitivity: ≤ 1.5%
3.10 Film density: >1.5
3.11 225KV H.V cable:5m,adopt flange to connect
3.12 ZLB-4 model Water cooling system
Water cooling system adopt anode forced cooling, the pipe for import and export is 10m.there are the protection switch of the water current relay and the temperature protection switch ,once the current over 4L/min the current relay will protect, than the protector will show alarm, once the temperature over 45°C,the X-Ray will show alarm.
3.13 T1016 model H.V Generator
H.V generator is self-governed cathode high-voltage generator,can output 225KV constant potential high-voltage,can meet the equipment under 225KV, make output allowance to make sure the stability and the security, it adopt high quality insulated oil and a double-voltage rectification circuit is adopted in the generator; together with the filament transformer make the x-ray tube producing a 0.1 –30 mA tube current.
3.14 T6000 model Controller
Tube voltage: precision ±1%,repeatitiveness superior ±1.5%, step≤1KV, actual tube voltage 40-225KV continuous adjust, steady voltage automatically.
Tube current: precision ±1%,repetitiveness superior ±1%, step≤0.1mA, actual tube current 0.1-10mA continuous adjust, steady current automatically.
T6000 Controller is adopted PLC control program which is the most advanced in the world at present, 5.7”big screen back light display board, Chinese interface, man-machine dialogue, dynamic display pattern, beautiful design, membrane switch, smooth surface display, easy to operate.
Controller screen displays current working parameters in digital mode, and all function parameters displayed on it can inter-lock diagnostic result.
The controller makes alarm in sound and light signals when the high voltage on and off. It also has the functions of safe interlock protect and alarm when water and electric conditions are not satisfied or in case wrong operations.
The controller has the functions of automatic warm-up, training and presetting the tube current and tube voltage. It makes the tube intelligent training automatically possible.
There are two operation modes, manual and automatic. The parameters of tube voltage, tube current, time preset, focus selection, inter-lock protect, man-machine dialogue and Chinese operation system are make sure no mistake operation. And the high voltage time accumulation, exposure fault memory, figures, sound and alarm prompt all are available.
It adopts high quality and advanced electronic parts. It has setting of many protect functions, system anti-interference abilities and error diagnose display function.
T6000 Controller can achieve the following functions:
--Control command sending such as high voltage on or off, etc.
--Training automatically control: It can identify whether it should execute training program. It can also cancel the function of training automatically.
--Identify faults and protect automatically: such as undercurrent, over-current, over-voltage, cooling problems, current protect, temperature protect etc.
--Timing or no-timing functions, 0-99minutes including no-timing mode
--The protect function of x-ray tube current and tube voltage up-limited through hardware setting and software setting.
3.15 Device for the X-ray tube moving and fixing
Mobile Constant Potential X-ray flaw Detectors(XYD-22510/3 model)